CVE-2025-0683

5.9 MEDIUM

📋 TL;DR

The Contec Health CMS8000 Patient Monitor transmits unencrypted patient data to a hard-coded public IP address when monitoring begins, potentially exposing sensitive health information. This affects healthcare facilities using these devices in their default configuration, creating privacy and compliance risks.

💻 Affected Systems

Products:
  • Contec Health CMS8000 Patient Monitor
Versions: All versions with default configuration
Operating Systems: Embedded medical device OS
Default Config Vulnerable: ⚠️ Yes
Notes: Vulnerability exists in default configuration when patient monitoring begins. Some reports suggest similar issues in Epsimed monitors.

⚠️ Manual Verification Required

This CVE does not have specific version information in our database, so automatic vulnerability detection cannot determine if your system is affected.

Why? The CVE database entry doesn't specify which versions are vulnerable (no version ranges provided by the vendor/NVD).

🔒 Custom verification scripts are available for registered users. Sign up free to download automated test scripts.

Recommended Actions:
  1. Review the CVE details at NVD
  2. Check vendor security advisories for your specific version
  3. Test if the vulnerability is exploitable in your environment
  4. Consider updating to the latest version as a precaution

⚠️ Risk & Real-World Impact

🔴

Worst Case

Patient health data is intercepted by malicious actors, leading to privacy violations, medical identity theft, or manipulation of critical health information during transmission.

🟠

Likely Case

Unauthorized access to patient monitoring data by entities controlling the hard-coded IP address, potentially violating HIPAA and other privacy regulations.

🟢

If Mitigated

Limited exposure if devices are isolated from external networks and data transmission is blocked, though default behavior remains risky.

🌐 Internet-Facing: HIGH - The hard-coded IP address is public, and devices may transmit data externally if network configurations allow it.
🏢 Internal Only: MEDIUM - Even on isolated networks, the transmission behavior occurs by default, though external exposure is prevented.

🎯 Exploit Status

Public PoC: ✅ No
Weaponized: LIKELY
Unauthenticated Exploit: ⚠️ Yes
Complexity: LOW

Exploitation requires network access to monitor traffic or control of the hard-coded IP. No authentication needed as transmission occurs automatically.

🛠️ Fix & Mitigation

✅ Official Fix

Patch Version: Not available

Vendor Advisory: https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01

Restart Required: No

Instructions:

No official patch exists. Follow FDA and CISA guidance for mitigation steps including network isolation and configuration changes.

🔧 Temporary Workarounds

Network Segmentation

all

Isolate patient monitors on dedicated VLANs with strict firewall rules blocking all external communication

Firewall Block Hard-Coded IP

all

Block all outbound traffic to the hard-coded IP address at network perimeter

🧯 If You Can't Patch

  • Disconnect devices from any network with internet access
  • Implement physical security controls to prevent unauthorized device access

🔍 How to Verify

Check if Vulnerable:

Monitor network traffic from CMS8000 devices for outbound connections to external IP addresses when patient monitoring begins

Check Version:

Check device firmware version through device interface or manufacturer documentation

Verify Fix Applied:

Confirm no outbound traffic reaches external networks and patient data transmission is contained within secure medical network

📡 Detection & Monitoring

Log Indicators:

  • Outbound connection attempts to external IP addresses from medical devices
  • Patient data transmission logs showing external destinations

Network Indicators:

  • TCP/UDP traffic from patient monitors to external IP addresses
  • Unencrypted medical data in network captures

SIEM Query:

source_ip IN (medical_device_ips) AND dest_ip IN (external_ips) AND protocol IN (tcp, udp)