CVE-2021-27408

7.5 HIGH

📋 TL;DR

This vulnerability allows attackers to read memory beyond intended boundaries in Welch Allyn medical device management tools, potentially leaking sensitive information. If chained with an out-of-bounds write vulnerability, it could lead to arbitrary code execution. Affected systems include Welch Allyn medical devices and management software across multiple product lines.

💻 Affected Systems

Products:
  • Welch Allyn Service Tool
  • Welch Allyn Connex Device Integration Suite - Network Connectivity Engine (NCE)
  • Welch Allyn Software Development Kit (SDK)
  • Welch Allyn Connex Central Station (CS)
  • Welch Allyn Service Monitor
  • Welch Allyn Connex Vital Signs Monitor (CVSM)
  • Welch Allyn Connex Integrated Wall System (CIWS)
  • Welch Allyn Connex Spot Monitor (CSM)
  • Welch Allyn Spot Vital Signs 4400 Device
  • Welch Allyn Spot 4400 Vital Signs Extended Care Device
Versions: Prior to specified patched versions for each product
Operating Systems: Not specified - embedded medical device systems
Default Config Vulnerable: ⚠️ Yes
Notes: Multiple Welch Allyn medical device products and management tools are affected across different version ranges.

📦 What is this software?

⚠️ Risk & Real-World Impact

🔴

Worst Case

Arbitrary code execution leading to complete system compromise, patient data theft, or disruption of medical device functionality.

🟠

Likely Case

Information disclosure exposing sensitive medical data or system information that could facilitate further attacks.

🟢

If Mitigated

Limited impact with proper network segmentation and access controls preventing exploitation attempts.

🌐 Internet-Facing: MEDIUM - Medical devices often have limited internet exposure but may be accessible through healthcare networks.
🏢 Internal Only: HIGH - Healthcare networks often have interconnected medical devices that could be targeted internally.

🎯 Exploit Status

Public PoC: ✅ No
Weaponized: UNKNOWN
Unauthenticated Exploit: ✅ No
Complexity: MEDIUM

Requires chaining with another vulnerability for code execution. Medical device vulnerabilities often have limited public exploit development due to specialized nature.

🛠️ Fix & Mitigation

✅ Official Fix

Patch Version: Welch Allyn Service Tool: v1.10+, Welch Allyn Connex NCE: v5.3+, Welch Allyn SDK: v3.2+, Welch Allyn CS: v1.8.6+, Welch Allyn Service Monitor: v1.7.0.0+, Welch Allyn CVSM: v2.43.02+, Welch Allyn CIWS: v2.43.02+, Welch Allyn CSM: v1.52+, Welch Allyn Spot 4400: v1.11.00+

Vendor Advisory: https://us-cert.cisa.gov/ics/advisories/icsma-21-152-01

Restart Required: Yes

Instructions:

1. Contact Welch Allyn for updated firmware/software. 2. Apply patches according to vendor instructions. 3. Restart affected devices. 4. Verify patch installation.

🔧 Temporary Workarounds

Network Segmentation

all

Isolate medical devices on separate network segments with strict access controls.

Access Control Restrictions

all

Implement strict authentication and authorization controls for medical device management interfaces.

🧯 If You Can't Patch

  • Implement strict network segmentation to isolate medical devices from general network traffic
  • Deploy intrusion detection systems monitoring for anomalous access to medical device management interfaces

🔍 How to Verify

Check if Vulnerable:

Check device/software version against affected versions list. Contact Welch Allyn support for vulnerability assessment tools.

Check Version:

Device/software specific - consult Welch Allyn documentation for version checking procedures.

Verify Fix Applied:

Verify installed version matches or exceeds patched versions listed in vendor advisory.

📡 Detection & Monitoring

Log Indicators:

  • Unusual access patterns to medical device management interfaces
  • Multiple failed authentication attempts
  • Unexpected device reboots or configuration changes

Network Indicators:

  • Anomalous network traffic to/from medical device management ports
  • Unexpected protocol usage on medical device interfaces

SIEM Query:

source_ip IN (medical_device_subnets) AND (event_type='authentication_failure' OR event_type='configuration_change')

🔗 References

📤 Share & Export